Device name:Dual–channel Portable calibrator
Device model:TPT-CAL-V200A
Device spec:100V/50A
Hardware function:
-Charge/Discharge current output calibration
-Charge/Discharge voltage output calibration
-Charge/Discharge power output calibration
-Charge/Discharge Current Meter Measurement Calibration
-Charge/Discharge Voltage Meter Measurement Calibration
-Charge/Discharge Current Output Verification
-Charge/Discharge Voltage Output Verification
-Charge/Discharge Power Output Verification
-Charge/Discharge Current Meter Measurement Verification
-Verification Report Generation
Device Description:
The 1-to-1 calibration is suitable for single-channel charge/discharge machines with fewer channels or portable single-channel charge/discharge machines.
Device Description:
The 1-to-1 calibration is suitable for single-channel charge/discharge machines with fewer channels or portable single-channel charge/discharge machines.
Supported Meter Models:
-Keysight 34401A;RS232
-Keysight 34461A;USB
-Keysight 34465A;USB
Communication wiring:
-ThinkLab Series Please refer to pages 10 of the hardware manual.
-ThinkP Series Please refer to pages 10 of the hardware manual.
-ThinkRP Series Please refer to pages 11 of the hardware manual.
Calibration wiring:
Please refer to pages 13 and 14 of the hardware manual.
Electrical Specifications:
Model | TPT-CAL-V200A |
Channel number | 1+1;Two-Channel Mutual Calibration Mode |
AC input | AC 100~240V/1A |
Applicable Specifications for Charge/Discharge Machines | 2~100V;0.001~50A |
Applicable Series for Charge/Discharge Machines | ThinkLab、ThinkP、ThinkRP |
Output Calibration and Verification | Current、Voltage;10% ~ 90%@F.S. |
Measurement Calibration and Verification | Current、Voltage;10% ~ 90%@F.S. & 25℃ |
Calibration and Verification Accuracy | Same as original model’s accuracy |
Verification Report Generation | Yes |
Operating Environment | 23±3℃;20 ~ 90 %HR |
Dimension (mm) | 220W x 84H x 355D |
This DM specification is for reference only and is subject to change without notice.